Highest signal per degree of the base-metal types — 68 µV/°C, to +900 °C.
What is Type E Thermocouple Wire?
Type E Thermocouple Wire is a base-metal thermocouple conductor pair, Chromel (NiCr 90/10) as the positive leg and Constantan (CuNi 55/45) as the negative leg, rated −200 … +900 °C with an EMF tolerance of ±1.5°C (0-275°C) / ±0.4% (275-900°C).
Author: RAS Thermal Engineering Team
Highest EMF output per degree (68 µV/°C), the best sensitivity among base metal thermocouples. Non-magnetic.
Limited high-temperature range compared to Type K/N. Not as widely adopted.
Compare all 5 types, ranges and tolerances →
Suitable markets: aerospace, laboratories, precision sensors and others
Technical Specifications
| Property | Specification |
|---|---|
| Diameter Range | φ0.025-5.0 mm |
| Tolerance | ±0.005mm (fine) / ±0.02mm (standard) |
| Temperature Range | −200 … +900 °C |
| Standards | IEC 60584-1 Class 1 / ASTM E230 / DIN EN 60584 |
| Constantan Grade | CuNi44 / WNX |
| Resistivity | ≈0.49 μΩ·m (CuNi 55/45) |
| Resistance per Metre | 0.8 mm 0.946-1.004 · 1.0 mm 0.606-0.642 · 1.2 mm 0.420-0.445 Ω/m |
| Tensile Strength | CuNi44 / WNX ≥420 N/mm² · TNX ≥390 N/mm² |
| Elongation | ≥20% |
| Typical Batch Test Data | CuNi44 1.0 mm, 2026-01-10: 0.617/0.608 Ω/m, 481/482 N/mm², elongation 32% |
| Surface Finish | Bright / Oxidised / Electro-polished |
| Packaging | DIN 125 / 160 / 250 / 355 spool / Coil / Custom reel |
| Coil Weight | 5kg / 10kg / 20kg standard or request |
| Traceability | Furnace batch number engraved on spool |
| Minimum Order | 5 kg bare wire · 10 kg stranded |
| Lead Time | 3-5 business days (stock 0.1-3.0 mm) · 2-3 weeks (non-stock) · 4-6 weeks (custom alloy) |
Available Product Forms
Bare Wire →
Solid single-strand, φ0.025-5.0 mm, bright or oxidised.
Stranded Wire →
Multi-strand for flexibility and vibration resistance. 7/19/37 strand.
How to specify Type E Thermocouple Wire
- Match the atmosphere and the expected continuous temperature to the limitation above before the type is fixed.
- Pick the form, Bare Wire or Stranded Wire, then read the conductor diameter and tolerance class off the specification table.
- Quote against IEC 60584-1 Class 1 and ask for the material test certificate carrying the furnace batch number.
Frequently Asked Questions (FAQ)
What is the sensitivity of Type E?
68 µV/°C, the highest EMF output per degree among base-metal thermocouples. A 10 °C change therefore moves the signal 0.68 mV, which is exactly what you want across a noisy plant floor.
What temperature range does Type E cover?
−200 °C to +900 °C, narrower at the top than Type K or N but with the largest signal per degree.
Are Type E conductors magnetic?
No. Both legs are non-magnetic, which suits induction-heating and other magnetic-field environments.
When should Type E be preferred over Type K?
Where signal strength matters more than ceiling temperature, such as electrically noisy plant and cryogenic instrumentation.
Bottom line: Type E trades ceiling temperature for signal — the highest output per degree of the base-metal types, with non-magnetic conductors. Choose it for electrically noisy or cryogenic measurement, not where the ceiling temperature decides.
References & Evidence
- IEC 60584-1:2013: Thermocouples, Part 1: EMF specifications and tolerances ↗
- ASTM E230/E230M: Temperature-Electromotive Force tables for standardized thermocouples ↗
- NIST ITS-90 Thermocouple Database: Type E reference table (SRD 60) ↗
- NIST ITS-90 Thermocouple Database: reference functions and downloadable tables (SRD 60) ↗
- NIST Monograph 175: Temperature-Electromotive Force Reference Functions and Tables ↗
- BIPM SI Brochure: definition of the kelvin and traceability of ITS-90 ↗